Home |
Search |
Today's Posts |
#71
![]() |
|||
|
|||
![]()
Roy Lewallen wrote:
There are several examples at Food for thought.txt available at http://eznec.com/misc/food_for_thought/. Regarding errors in the first food_for_thought: A 100w source equipped with a circulator and load while looking into an open line, will generate 100w and dissipate 100w in the circulator load. That 100w is definitely not free power. It can be demonstrated to have made a round trip to the open end of the feedline and then back to the circulator load. The error in your thinking is that the source would see an open circuit when it is equipped with a circulator and load. It won't. It will *always* see the Z0 of the feedline as its load (assuming the circulator load equals Z0). That's the purpose of using the circulator and load - to allow the source to see a fixed load equal to Z0. -- 73, Cecil http://www.qsl.net/w5dxp ----== Posted via Newsfeeds.Com - Unlimited-Uncensored-Secure Usenet News==---- http://www.newsfeeds.com The #1 Newsgroup Service in the World! 120,000+ Newsgroups ----= East and West-Coast Server Farms - Total Privacy via Encryption =---- |
#72
![]() |
|||
|
|||
![]()
Richard Clark wrote:
Cecil Moore wrote: Optics engineers figured it out a long time ago. And you have consistently failed in its demonstration - so what? I can lead you to water but I can't make you drink. -- 73, Cecil http://www.qsl.net/w5dxp ----== Posted via Newsfeeds.Com - Unlimited-Uncensored-Secure Usenet News==---- http://www.newsfeeds.com The #1 Newsgroup Service in the World! 120,000+ Newsgroups ----= East and West-Coast Server Farms - Total Privacy via Encryption =---- |
#73
![]() |
|||
|
|||
![]()
On Wed, 29 Jun 2005 22:39:10 -0500, Cecil Moore
wrote in : Frank Gilliland wrote: There lies our misperceptions; I was not referring to using an HF SWR meter designed for coax and plugging it into 450 ohm ladder line. But I specifically stated above the Z0 environment was different from 50 ohms. The same type of error happens when one uses a 50 ohm SWR meter in a 75 ohm coaxial line. If that were true then the mere existence of standing waves could render any measurements worthless. Regardless, I did the experiment a long time ago -- take a 50 ohm SWR meter and plug it into a 75 ohm line -- it gives you almost the same measurement (in fact, I didn't see -any- difference at all). Any small error you might see is, as I said before, insignificant, especially considering the reason you are measuring SWR in the first place. The objective is simply to get the reading as low as practially possible. If you feel the need to quibble about a couple tenths of a point on a ratio then maybe you're spending a little too much time playing with the calculator instead of the antenna. ----== Posted via Newsfeeds.Com - Unlimited-Uncensored-Secure Usenet News==---- http://www.newsfeeds.com The #1 Newsgroup Service in the World! 120,000+ Newsgroups ----= East and West-Coast Server Farms - Total Privacy via Encryption =---- |
#74
![]() |
|||
|
|||
![]()
On Wed, 29 Jun 2005 23:10:25 -0500, Cecil Moore
wrote in : Frank Gilliland wrote: Wes Stewart wrote: In a word, baloney. The error is independent of length. A zero length bridge calibrated at 75 ohm is in error when measuring in a 50 ohm system. Period. Prove it. A 75 ohm bridge is expecting the ratio of voltage to current to be 75 for a matched system. In a 50 ohm matched system, the ratio of voltage to current will be 50. Therefore, the 75 ohm bridge won't be balanced. A 50 ohm bridge would be balanced. The bridge is calibrated to the impedance of the directional coupler (which is usually built to match the expected line impedance, but cannot be "zero length" in the present state of reality). If the impedance of the signal is different than what is expected by the bridge then your power measurements will probably be wrong (to what extent they are wrong may or may not be important). But if that's the case then any error will be the same by percentage and sign for both forward =AND= reflected power because the impedance of the signal is the same for both forward and reflected power. IOW, the ratio is the same -despite- the impedance. If you don't believe me, try it yourself. ----== Posted via Newsfeeds.Com - Unlimited-Uncensored-Secure Usenet News==---- http://www.newsfeeds.com The #1 Newsgroup Service in the World! 120,000+ Newsgroups ----= East and West-Coast Server Farms - Total Privacy via Encryption =---- |
#75
![]() |
|||
|
|||
![]()
On Wed, 29 Jun 2005 22:53:24 -0700, Frank Gilliland
wrote: cannot be "zero length" in the present state of reality). If the impedance of the signal is different than what is expected by the bridge then your power measurements will probably be wrong (to what extent they are wrong may or may not be important). But if that's the case then any error will be the same by percentage and sign for both forward =AND= reflected power because the impedance of the signal is the same for both forward and reflected power. IOW, the ratio is the same -despite- the impedance. Lets make an assumption that we are talking about lossless lines. (If you are not, then the reflectometer does not provide an accurate indication of forward and reverse power.) If you use an ideal 50 ohm reflectometer (that means it is a negligibly short 50 ohm through line and it is nulled to show zero reflected power when connected to a 50+j0 load) to measure conditions in a line, the power flow at that point is the indicated Pf-Pr. If you had placed an ideal 75 ohm instrument in that spot, the readings are not necessarily in the same ratio (they are unlikely to be so), but the difference between Pf and Pr will be the same. The only other inference you can make from one instrument with regard to the other will be if one of the instruments shows zero reflected power, then you know the VSWR that the other instrument will indicate. Real instruments aren't of zero length, but some types of design are so close to it at low HF frequencies, you will not detect the error that is introduced. Owen -- |
#76
![]() |
|||
|
|||
![]() Updated: On Wed, 29 Jun 2005 22:53:24 -0700, Frank Gilliland wrote: cannot be "zero length" in the present state of reality). If the impedance of the signal is different than what is expected by the bridge then your power measurements will probably be wrong (to what extent they are wrong may or may not be important). But if that's the case then any error will be the same by percentage and sign for both forward =AND= reflected power because the impedance of the signal is the same for both forward and reflected power. IOW, the ratio is the same -despite- the impedance. Lets make an assumption that we are talking about distortionless lines. (If you are not, then the reflectometer does not provide an accurate indication of forward and reverse power.) If you use an ideal 50 ohm reflectometer (that means it is a negligibly short 50 ohm through line and it is nulled to show zero reflected power when connected to a 50+j0 load) to measure conditions in a line, the power flow at that point is the indicated Pf-Pr. If you had placed an ideal 75 ohm instrument in that spot, the readings are not necessarily in the same ratio (they are unlikely to be so), but the difference between Pf and Pr will be the same. The only other inference you can make from one instrument with regard to the other will be if one of the instruments shows zero reflected power, then you know the VSWR that the other instrument will indicate. Real instruments aren't of zero length, but some types of design are so close to it at low HF frequencies, you will not detect the error that is introduced. Owen -- |
#77
![]() |
|||
|
|||
![]()
I'm not quite sure what you are trying to say Frank.
Frank Gilliland wrote: On Wed, 29 Jun 2005 23:10:25 -0500, Cecil Moore wrote in : The bridge is calibrated to the impedance of the directional coupler (which is usually built to match the expected line impedance, but cannot be "zero length" in the present state of reality). The direction coupler samples voltage across and current through a given point. There is always a current transformer of some type and a voltage sample through some type of divider. The "voltages" representing E and I are summed before detection (conversion to dc). The "directivity" comes because the current phase sample is reversed 180 degrees from the summing phase, causing voltages to subtract. This means the directional coupler is calibrated for a certain ratio of voltage and current, so when they exist you have twice the voltage in the direction where E and I add, and zero voltage where they subtract. If the impedance of the signal is different than what is expected by the bridge then your power measurements will probably be wrong (to what extent they are wrong may or may not be important). But if that's the case then any error will be the same by percentage and sign for both forward =AND= reflected power because the impedance of the signal is the same for both forward and reflected power. IOW, the ratio is the same -despite- the impedance. ?What does that mean? If the directional coupler is calibrated at 50 ohms and you use it in a 75 ohm system you won't get a total reflected null even if the 75 ohm line has a 1:1 SWR. But if you subtract reflected power from forward power readings you will get the correct power, within linearity and calibration limits of the "meter system". This has nothing to do with standing waves. It has only to do with the relationship between current and voltage at the point where the directional coupler is inserted. I'm not sure if you are saying that or not. 73 Tom |
#78
![]() |
|||
|
|||
![]()
Frank Gilliland wrote:
Cecil Moore wrote: But I specifically stated above the Z0 environment was different from 50 ohms. The same type of error happens when one uses a 50 ohm SWR meter in a 75 ohm coaxial line. If that were true then the mere existence of standing waves could render any measurements worthless. Regardless, I did the experiment a long time ago -- take a 50 ohm SWR meter and plug it into a 75 ohm line -- it gives you almost the same measurement (in fact, I didn't see -any- difference at all). Please run it again in the following configuration: Xmtr--1/4WL 75 ohm line--SWR meter--1/4WL 75 ohm line--50 ohm load The SWR meter will read 2.25:1 when the actual SWR is 1.5:1 Xmtr--1/2WL 75 ohm line--SWR meter--1/2WL 75 ohm line--50 ohm load The SWR meter will read 1:1 when the actual SWR is 1.5:1 Any small error you might see is, as I said before, insignificant, especially considering the reason you are measuring SWR in the first place. A 50% error in SWR reading is NOT insignificant. The objective is simply to get the reading as low as practially possible. If you feel the need to quibble about a couple tenths of a point on a ratio then maybe you're spending a little too much time playing with the calculator instead of the antenna. A 50% error in SWR is NOT a couple tenths of a point. -- 73, Cecil http://www.qsl.net/w5dxp ----== Posted via Newsfeeds.Com - Unlimited-Uncensored-Secure Usenet News==---- http://www.newsfeeds.com The #1 Newsgroup Service in the World! 120,000+ Newsgroups ----= East and West-Coast Server Farms - Total Privacy via Encryption =---- |
#79
![]() |
|||
|
|||
![]()
Frank Gilliland wrote:
Cecil Moore wrote: A 75 ohm bridge is expecting the ratio of voltage to current to be 75 for a matched system. In a 50 ohm matched system, the ratio of voltage to current will be 50. Therefore, the 75 ohm bridge won't be balanced. A 50 ohm bridge would be balanced. The bridge is calibrated to the impedance of the directional coupler (which is usually built to match the expected line impedance, but cannot be "zero length" in the present state of reality). If the impedance of the signal is different than what is expected by the bridge then your power measurements will probably be wrong (to what extent they are wrong may or may not be important). But if that's the case then any error will be the same by percentage and sign for both forward =AND= reflected power because the impedance of the signal is the same for both forward and reflected power. IOW, the ratio is the same -despite- the impedance. The error is NOT the same percentage. In a matched 50 ohm system, the 75 ohm bridge reflected power reading will be off by an infinite percentage, i.e. division by zero. If you don't believe me, try it yourself. I have tried it and you are wrong. Maybe you should try it. -- 73, Cecil http://www.qsl.net/w5dxp ----== Posted via Newsfeeds.Com - Unlimited-Uncensored-Secure Usenet News==---- http://www.newsfeeds.com The #1 Newsgroup Service in the World! 120,000+ Newsgroups ----= East and West-Coast Server Farms - Total Privacy via Encryption =---- |
#80
![]() |
|||
|
|||
![]()
Owen wrote:
Real instruments aren't of zero length, but some types of design are so close to it at low HF frequencies, you will not detect the error that is introduced. The error that we are talking about has nothing to do with the length of the directional coupler. The error that we are talking about has everything to do with an infinite error in the measurement of reflected power. Infinite errors are hard to sweep under the rug. -- 73, Cecil http://www.qsl.net/w5dxp ----== Posted via Newsfeeds.Com - Unlimited-Uncensored-Secure Usenet News==---- http://www.newsfeeds.com The #1 Newsgroup Service in the World! 120,000+ Newsgroups ----= East and West-Coast Server Farms - Total Privacy via Encryption =---- |
Reply |
Thread Tools | Search this Thread |
Display Modes | |
|
|