Home |
Search |
Today's Posts |
#1
![]() |
|||
|
|||
![]()
Well since I am boxing the loose tubes I have, guess I had better test
them! I have a military I-177 Made by Daven that sez that the gm of the tubes is 3x higher than it should be. A 6K6 should be 1600 but I read 4200!!! Pins the needle on the 3000 range and reads 4200 on the 6000 range. All known good tubes tested read high. I do not see a pot or adjustable power resistor to change the calibration. Looked @ the calibration info on the Hickok 600 series, but is too different to apply to the I-177. Any ideas? Anyone have the calibration routine for the I-177? The mil manuals I have found do not have a calibration routine. Thanks Randy. |
#2
![]() |
|||
|
|||
![]()
"AB9GO" ha scritto nel messaggio oups.com...
Well since I am boxing the loose tubes I have, guess I had better test them! I have a military I-177 Made by Daven that sez that the gm of the tubes is 3x higher than it should be. A 6K6 should be 1600 but I read 4200!!! Pins the needle on the 3000 range and reads 4200 on the 6000 range. All known good tubes tested read high. I do not see a pot or adjustable power resistor to change the calibration. Looked @ the calibration info on the Hickok 600 series, but is too different to apply to the I-177. Any ideas? Anyone have the calibration routine for the I-177? The mil manuals I have found do not have a calibration routine. Thanks Randy. I do not know which is the story of your tube tester. A possibility is that somone replaced the two wire-wound potentiometers with others of different values (it happened to me!). It is VERY difficult to find a replacement for them. One is a 3000-ohm potentiometer having a non-linear taper, the other one consists of two ganged low-value potentiomers. I would check them with an ohmeter before proceeding to isolate the fault. 73 Tony I0JX |
#3
![]() |
|||
|
|||
![]()
Randall, the I-177 tube tester uses 3 shunts for 3 ranges of gm measure.
My I-177 was defective, the shunts were broken, so the measure was not ok. These shunts are near a switch, as I can remember. The repair was done a lot of years ago, so control by yourself ! bye from Piero ( ex I1SPO, ex I5SPO). "AB9GO" ha scritto nel messaggio oups.com... Well since I am boxing the loose tubes I have, guess I had better test them! I have a military I-177 Made by Daven that sez that the gm of the tubes is 3x higher than it should be. A 6K6 should be 1600 but I read 4200!!! Pins the needle on the 3000 range and reads 4200 on the 6000 range. All known good tubes tested read high. I do not see a pot or adjustable power resistor to change the calibration. Looked @ the calibration info on the Hickok 600 series, but is too different to apply to the I-177. Any ideas? Anyone have the calibration routine for the I-177? The mil manuals I have found do not have a calibration routine. Thanks Randy. |
Reply |
Thread Tools | Search this Thread |
Display Modes | |
|
|
![]() |
||||
Thread | Forum | |||
F.S. EICO 667 Tube Tester | Equipment | |||
F.S. EICO 667 Tube Tester | Equipment | |||
Tube Tester | Boatanchors | |||
Tube Tester | Boatanchors | |||
WTB: Jackson 648-1T Tube Tester Paper | Boatanchors |